Melamine ELISA Test Kit
Catalog No. LSY-10026
1. Principle
This test kit is based on the indirect competitive enzyme immunoassay for the detection of Melamine in the sample. The coupling antigen is pre-coated on the micro-well stripes. The Melamine in the sample and the coupling antigens pre-coated on the micro-well stripes compete for anti-Melamine antibodies. After the addition of the enzyme conjugate, the TMB substrate is added for coloration. The optical density (OD) value of the testing sample has a negative correlation with the Melamine concentration in the sample. This value is compared to the standard curve and the Melamine concentration is subsequently obtained.
2. Technical specifications
Sensitivity: 0.5ppb
Incubator temperature: 25℃
Incubator time: 30min~15min
Detection limit:
Fresh milk, pure milk 5ppb
Yogurt, milk powder 10ppb
Milk chocolate, cheese, butter, cream 5ppb
Cross-reaction rate:
Melamine 100%
Recovery rate:
Fresh milk, pure milk 90±25%
Yogurt, milk powder 85±25%
Milk chocolate, cheese, butter, cream 85±25%
3. Components
1 | Micro-well strips | 12 strips with 8 removable wells each | |
2 | 6× standard solution (1 mL each) | 0ppb | 0.5ppb |
1.5ppb | 4.5ppb | ||
13.5ppb | 40.5ppb | ||
3 | Enzyme conjugate | 7ml | red cap |
4 | Antibody working solution | 7ml | blue cap |
5 | Substrate A | 7ml | white cap |
6 | Substrate B | 7ml | black cap |
7 | Stop solution | 7ml | yellow cap |
8 | 20× concentrated washing buffer | 15ml | white cap |
9 | 20x sample extracting solution | 50ml | transparent cap |
10 | 10x sample dilution | 50ml | blue cap |
4. Materials required but not provided
1) Equipments: microplate reader, printer, vortex, centrifuge, measuring pipets, balance( a reciprocal sensibility of 0.01 g).
2) Micropipettors: single-channel 20~200 µL, 100~1000 µL; and multi-channel 30~300 μl;
5. Storage and expiry date
Storage: stored at 2-8 ℃, not frozen.
Expiry date: 12 months; date of production is on the box
Note: If the Vacuum package of microplate has leakage, it is still valid to use, do not affect the test result, be relax to use.