Read More

Porcine Circovirus (PCV2) ELISA test Kit for swine

  • Sep 09, 2026   Anon

 Porcine Circovirus (PCV2) ELISA test Kit

Catalog No. LSY-30007

1. Usage

The GreenSpring® porcine circovirous(PCV2) ELISA Test kit is used for detection of porcine circovirus type 2 antibody in porcine serum; assessment of immunity conditions against porcine circovirus in the pig farm and investigation of the epidemiology of the porcine circovirus.While the result can not distinguish immune antibody and infectious antibody, to confirm that is antibody by vaccine or wild virus infection needs to combine clinical data, immune status, and other diagnostic methods.

2. Principle

The GreenSpring® PCV2 antibody ELISA test kit uses the principle of solid-phase enzyme-linked immunosorbent assay (ELISA), and consists of a microwell reaction plate coated with high-purity PCV2 antigen, anti-pig IgG labeled with horseradish peroxidase, and other reagents. The reaction mechanism is that the coated antigen is combined with the PCV-Ab in the sample, and then combined with the enzyme-labeled anti-pig IgG antibody to form a "coated antigen+PCV-Ab+enzyme-labeled anti-pig IgG antibody" complex. Add adding Substrate, Catalytic reaction color development. The color depth is directly proportional to the amount of PCV-Ab. When the color reaction of the sample is detected, the result obtained by the microplate reader exceeds the set threshold, and the result is positive, which indicates that the immune has produced antibodies or there is a natural infection.

3. The kit components 

1

PCV antigen coated microplate

96T X 2

2

Enzyme conjugate

22ml

yellow lid

3

Sample diluent solution

50ml

transparent lid

4

PCV-IgG Negative control serum

1.5ml

green lid

5

PCV-IgG Positive control serum

1.5ml

red lid

6

Substrate

12ml X2

orange lid

7

Stop solution

12ml

blue lid

8

1concentrated washing buffer

50ml

white lid

9

Adhesive Foil

2 pieces


10

Instruction

1 piece


 

4. Materials Required But Not Provided

1 Microplate Reader (double-wave length: 450/630 nm).

2 Precise micropipette (single-channel 1-100ul0.5-10ulmulti-channel 30-300ul)

3 Constant temperature box or water bath box.

4 Oscillator.

5 Disposable tips (10ul, 200ul)

       6 Deionized water 

5. Precautions and warnings for users

      1. Read the Manual carefully before use.

      2. Do not use reagents expired, do not mix reagents from different lots.

3. Experiment rubbish should be dealt with high pressure steam sterilization at 121 for 30 minutes, or treated with 5.0g/L sodium hypochlorite disinfectant for 30 minutes, then discard.

  4. MicroWell plate removed from the refrigerated environment should be balanced moisture to dry at room temperature, then can be opened. Put back unused MicroWell plate into dry foil bag and sealed at 4 . Unused liquid reagent should cover caps, store at 2-8 in dark with other group components.

      5. Should use Micropipettor to add sample and reagents, and often proof its accuracy.

6. When adding washing buffer, should be full but no overflow, avoid appearing free enzyme at mouth of well or cross pollution between wells.

7. Stop solution is corrosive, use large amount of water to wash immediately when touch the skin or clothes.

 

Specifications: 96 wells×2.

Expiry date: 12 months.

Storage: Store at 2~8, in the dark.

Production Date: On outer-packing of the test kit.

 

Shenzhen Lvshiyuan Biotechnology Co., Ltd

D Building, National Biological Industrial Park of Marinelife, No.2 Binhai Road, Dapeng, Shenzhen, 518120 China

Tel. 86-755-28438788/86-18165709090

Fax 86-755-28938800

Email:  info@lsybt.com 

www.lsybt.com